Chapter 5
5-6
V60 Ventilator Service Manual
1049766 Rev A
Diagnostic Mode and Troubleshooting
5.1.3
Restoring Default Settings
The Restore Default Settings screen allows you to restore ventilation and alarm
settings to their default values. This function does
not
change other settings
such as brightness, loudness, language, Baud rate, or pressure units.
1. From the System Settings screen, touch
Restore Default Settings
.
2. Touch
Restore Defaults
on the Restore Default Settings screen (Figure
5-6).
To exit without restoring default settings, touch
Cancel
.
Figure 5-6: Restore Default Settings Screen
3. When all ventilator and alarm settings have been restored to factory
default values, the screen displays this message:
Ventilation and Alarm Settings have been restored to factory defaults.
Содержание Respironics V60
Страница 1: ...Respironics V60 Ventilator Service Manual ...
Страница 8: ...viii V60 Ventilator Service Manual 1049766 Rev A Table of Contents This page is intentionally blank ...
Страница 204: ...Chapter 8 8 50 V60 Ventilator Service Manual 1049766 Rev A Performance Verification This page is intentionally blank ...
Страница 220: ...Chapter 9 9 16 V60 Ventilator Service Manual 1049766 Rev A Replacement Parts List This page is intentionally blank ...
Страница 242: ...Appendix A A 14 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...
Страница 244: ...Appendix B B 2 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...