1049766 Rev A
V60 Ventilator Service Manual
8-1
Chapter 8. Performance Verification
Performance verification verifies the integrity of the sensors and other critical
components in the ventilator using external measurement devices. The
performance verification consists of several tests (Table 8-1). The type of
service performed determines which tests are required (Table 8-2). Run all
tests in order.
CAUTION:
Performance verification is to be performed only by a qualified service
technician.
Table 8-1: Performance Verification Tests
Test number
Description
1
Electrical safety
2
Leak tests
3
Ventilator controls
4
Pressure accuracy
5
Air delivery/flow accuracy*
6
Oxygen flow accuracy*
7
Oxygen accuracy
8
S/T performance
9
Alarms
10
Power fail
11
Internal battery
* Flow testing above 1951 m (6400 ft.) cannot attain 130 SLPM.