Chapter 10
10-8
V60 Ventilator Service Manual
1049766 Rev A
Specifications
10.10 Other
specifications
10.11 Contact
Information
For further information or technical assistance, contact Respironics Customer
Service:
•
By email:
•
In the U.S.:
800-345-6443
•
Outside the U.S.:
724-387-4000
(phone) or
724-387-5012
(fax)
Table 10-11: Other specifications
Parameter
Specification
Flow delivery
150 L/min at 40 cmH
2
O at 1951 m (6400 ft) altitude
(10% degradation in flow at 2286 m/7500 ft)
Demand flow range
-240 to 240 L/min BTPS
Pressure range
4 to 40 cmH
2
O
Dynamic pressure regulation
± (2 cmH
2
O + 4% of target)
NOTE:
Negative (subatmospheric) pressure
settings are not available.
Start-up time
Ready to ventilate 9 seconds after power on
Triggering, cycling, and leak tol-
erance
As per the Digital Auto-Trak
Sensitivity algorithms (see
V60 Ventilator Operator’s Manual
)
Inspiratory and expiratory pres-
sure drop: measured at patient
connection, when the recom-
mended breathing system is in
use and normal ventilation is
compromised by the total or par-
tial loss of power supply
< 40 cmH
2
O (at 60 LPM)
< 1.5 cmH
2
O (at 30 LPM)
Audio alarm loudness
60 to 95 dBA (primary alarm)
≥
65 dBA (backup alarm)
Acoustic noise
Less than 45 dB (A) at 1 m)
Содержание Respironics V60
Страница 1: ...Respironics V60 Ventilator Service Manual ...
Страница 8: ...viii V60 Ventilator Service Manual 1049766 Rev A Table of Contents This page is intentionally blank ...
Страница 204: ...Chapter 8 8 50 V60 Ventilator Service Manual 1049766 Rev A Performance Verification This page is intentionally blank ...
Страница 220: ...Chapter 9 9 16 V60 Ventilator Service Manual 1049766 Rev A Replacement Parts List This page is intentionally blank ...
Страница 242: ...Appendix A A 14 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...
Страница 244: ...Appendix B B 2 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...