Chapter 4
Periodic Maintenance
1049766 Rev A
V60 Ventilator Service Manual
4-3
3. Touch
Service
to display the V60 Ventilator Information screen (Figure
4-2).
Figure 4-2: Ventilator Information Screen
4. Record the following information on the Performance Verification Data
Form (Chapter 8):
•
Software options
•
Ventilator serial number
•
Software version
•
Total power-on hours
•
Hours since last PM
•
Battery lot number
•
Battery manufacture date
5. Generate a diagnostic report (DRPTA) for the ventilator (Chapter 6).
6. Turn off the V60 by pressing the
ON/Shutdown
button.
7. Disconnect all power, remove the top cover, and remove the EMI
shroud from the power supply (section 7.16).
8. Use an electrostatic discharge (ESD) safe vacuum to remove any
accumulated dust and debris from the inside the ventilator enclosure.
9. Inspect all electrical and pneumatic connections.