Chapter 7
Component Removal/Installation
1049766 Rev A
V60 Ventilator Service Manual
7-27
7.24 Air and Oxygen
Flow Sensor
Assembly
Follow these steps to remove the air and oxygen flow sensor assembly
(Figure 7-25). Reverse to install.
1. Remove GDS (section 7.20).
2. Disconnect the flow sensor cable from the DA PCBA and flow sensor
assembly.
3. Remove the screws that attach the DA PCBA to the flow sensor
assembly.
4. Remove screws that attach the flow sensor assembly to the manifold
block.
NOTE:
When installing a new flow sensor assembly, install the new o-rings.
Figure 7-25: Removing the Flow Sensor Assembly
Flow sensor
ribbon cable
DA PCBA
GDS
Flow sensor
assembly
M3 x 8 screw (x2)
(62 + 5.7 N cm/
5.5 + 0.5 in.-lbf)
M3 washer (x2)
DA to MC
ribbon
cable
connector
M3 x 14 screw (x4)
(62 + 5.7 N cm/
5.5 + 0.5 in.-lbf)
DA PCBA