Chapter 8
8-30
V60 Ventilator Service Manual
1049766 Rev A
Performance Verification
3. Select these ventilator settings and alarm limits:
4. With the oxygen source disconnected, touch
O2
and set to 22%.
5. Verify that the ventilator alarms and displays
Oxygen Not Available
.
6. Touch
O2
and set to 21%, reset alarms.
7. Connect an oxygen source to the ventilator.
NOTE:
Calibrate the oxygen analyzers sensor prior to proceeding.
8. Verify that the set O
2
% values corresponds to the measured values:
9. Touch
O2
and set to 21%.
10. The oxygen accuracy test is complete.
Ventilator settings
Alarm limits
Mode: S/T
High Rate: 30 BPM
IPAP: 20 cmH
2
O
Low Rate: 4 BPM
EPAP: 4 cmH
2
O
Hi Vt: 2500 mL
Rate: 12 BPM
Lo Vt: OFF mL
I-Time: 1.00 sec
HIP: 50 cmH
2
O
Rise: 1
LIP: OFF cmH
2
O
Ramp: OFF
Low VE: OFF L/min
O2: 21%
LIP T: 20 sec
Touch
O2
and
set to:
Verify that analyzer measures:
21%
18 to 24% (after 12 breaths)
30%
27 to 33% (after 12 breaths)
60%
57 to 63% (after 12 breaths)
100%
97 to 103% (after 12 breaths)
Содержание Respironics V60
Страница 1: ...Respironics V60 Ventilator Service Manual ...
Страница 8: ...viii V60 Ventilator Service Manual 1049766 Rev A Table of Contents This page is intentionally blank ...
Страница 204: ...Chapter 8 8 50 V60 Ventilator Service Manual 1049766 Rev A Performance Verification This page is intentionally blank ...
Страница 220: ...Chapter 9 9 16 V60 Ventilator Service Manual 1049766 Rev A Replacement Parts List This page is intentionally blank ...
Страница 242: ...Appendix A A 14 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...
Страница 244: ...Appendix B B 2 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...