Chapter 8
8-32
V60 Ventilator Service Manual
1049766 Rev A
Performance Verification
5. Select these ventilator settings and alarm limits, and reset any alarms
if necessary:
6. Allow the ventilator to operate for at least one minute.
7. Watch the waveforms on the V60 display to verify that the unit cycles
between IPAP and EPAP.
8. Verify that the set IPAP and EPAP values correspond to these
measurements:
9. The S/T performance test is complete.
Ventilator settings
Alarm limits
Mode: S/T
High Rate: 30 BPM
IPAP: 15 cmH
2
O
Low Rate: 4 BPM
EPAP: 4 cmH
2
O
Hi Vt: 2500 mL
Rate: 12 BPM
Lo Vt: OFF mL
I-Time: 1.00 sec
HIP: 50 cmH
2
O
Rise: 2
LIP: OFF cmH
2
O
Ramp: OFF
Low VE: OFF L/min
O2: 21%
LIP T: 20 sec
Select this setting:
Verify that analyzer measures:
IPAP = 40 cmH
2
O
36.4 to 43.6 cmH
2
O (PIP)
EPAP = 25 cmH
2
O
22 to 28 cmH
2
O (PEEP)