Chapter 2
Warnings and Cautions
1049766 Rev A
V60 Ventilator Service Manual
2-3
WARNING:
The nurse call/remote alarm should be considered a backup to the ventilator’s
primary alarm system.
WARNING:
To ensure that the alarm will be heard, make sure the alarm loudness is
adequate and avoid blocking the alarm speakers beneath the ventilator.
CAUTION:
Federal law (USA) restricts this device to sale by or on the order of a
physician.
CAUTION:
The Respironics V60 ventilator is designed to operate in the temperature
range of 5 to 40 ºC (41 to 104 ºF). To minimize the risk of overheating the
device, do not operate adjacent to heaters or other heat sources.
NOTE:
The displays shown in this manual may not exactly match what you see on
your own ventilator.
NOTE:
Pressures are indicated on the ventilator in cmH
2
O. Millibars and
hectopascals (hPa) are used by some institutions instead. Since 1
millibar equals 1 hPa, which equals 1.016 cmH
2
O, the units may be
used interchangeably.
NOTE:
The ventilator is not intended for use as an ambulance transport ventilator
or as an Automatic Transport Ventilator as described by the American
Hospital Association and referenced by the FDA. It is intended to allow
the patient to be transported within the hospital setting using a cart to
move the ventilator.
NOTE:
When attachments or other components or subassemblies are added to
the ventilator breathing system, the pressure gradient across the
ventilator breathing system, measured with respect to the ventilator
outlet, may increase.
NOTE:
The Respironics V60 ventilator parts that have patient contact are free of
latex.
NOTE:
If an alarm persists for no apparent reason, discontinue ventilator use and
contact Respironics.
NOTE:
If you detect any unexplained changes in the performance or visual
displays of the ventilator, discontinue ventilator use and contact
Respironics.
NOTE:
The Respironics V60 ventilator does not support automatic record
keeping.
Содержание Respironics V60
Страница 1: ...Respironics V60 Ventilator Service Manual ...
Страница 8: ...viii V60 Ventilator Service Manual 1049766 Rev A Table of Contents This page is intentionally blank ...
Страница 204: ...Chapter 8 8 50 V60 Ventilator Service Manual 1049766 Rev A Performance Verification This page is intentionally blank ...
Страница 220: ...Chapter 9 9 16 V60 Ventilator Service Manual 1049766 Rev A Replacement Parts List This page is intentionally blank ...
Страница 242: ...Appendix A A 14 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...
Страница 244: ...Appendix B B 2 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...