Chapter 7
Component Removal/Installation
1049766 Rev A
V60 Ventilator Service Manual
7-3
7.2 Air Inlet Filter
Follow these steps to remove the air inlet filter (Figure 7-2). Reverse to install.
1. Turn the unit OFF and disconnect the power cord from the outlet.
2. Twist the locking pin 1/4 turn with a slotted screwdriver or by hand,
and pull down on locking pin to remove right side panel (locking pin is
captive).
3. Remove air inlet filter.
Figure 7-2: Removing the Air Inlet Filter
7.3 Internal Battery
Follow these steps to remove and install the internal battery (Figure 7-3).
1. Disconnect power (section 7.1).
2. Disconnect the internal battery from the power harness connector.
3. Remove the side panel by turning the captive Phillips head fastener a
1/4 turn and releasing.
4. Using a 3-mm hex wrench, remove the battery bracket by removing
two screws.
Right side
panel
Air inlet
filter