Chapter 7
7-6
V60 Ventilator Service Manual
1049766 Rev A
Component Removal/Installation
7.5 AC Inlet
Follow these steps to remove the AC inlet (Figure 7-5). Reverse to install.
1. Remove top cover (section 7.4).
2. Remove EMI shroud from power supply (section 7.16).
3. Remove the screws that attach the power cord retainer, then remove
retainer and power cord.
4. Remove screws that attach the AC inlet to the back panel.
5. Disconnect AC inlet cables from power supply and ground terminal.
6. Feed cable through back panel cutout, and pull AC inlet from back
panel.
CAUTION:
Install ground wires and nuts in this order to ensure proper grounding:
1. Kep nut
2. AC inlet wire
3. Kep nut
4. Power supply connector wire
5. GDS ground wire
6. Kep nut
NOTE:
When reinstalling, orient the AC inlet with ground terminal as shown.
Figure 7-5: Removing AC Inlet
AC inlet
AC inlet wire to
ground terminal
M2.5 x 8 screw (x2)
(34 + 3.4 N cm/
3 + 0.3 in.-lbf)
Power cord
retainer
Power
cord
AC inlet wires to
power supply
Ground terminal
M4 x 8 screw (x2)
(113 + 11.3 N cm/
10 + 1 in.-lbf)
M4 kep nut (x3)
(113 + 11.3 N cm/
10 + 1 in.-lbf)
Содержание Respironics V60
Страница 1: ...Respironics V60 Ventilator Service Manual ...
Страница 8: ...viii V60 Ventilator Service Manual 1049766 Rev A Table of Contents This page is intentionally blank ...
Страница 204: ...Chapter 8 8 50 V60 Ventilator Service Manual 1049766 Rev A Performance Verification This page is intentionally blank ...
Страница 220: ...Chapter 9 9 16 V60 Ventilator Service Manual 1049766 Rev A Replacement Parts List This page is intentionally blank ...
Страница 242: ...Appendix A A 14 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...
Страница 244: ...Appendix B B 2 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...