Chapter 8
8-42
V60 Ventilator Service Manual
1049766 Rev A
Performance Verification
8.7.9
Test 9: Alarms
8.7.10
Test 10: Power Fail
Symptom
Recommended Repair
Normally open or
normally closed
relay test failure
(continuity check)
1. Verify continuity of remote alarm test cable.
2. Replace CPU PCBA.
Primary alarm
volume not
adjustable
Slave in PCBAs in this order until the problem is
resolved: MC PCBA, PM PCBA, CPU PCBA.
Symptom
Recommended Repair
Backup audible
alarm fails to
sound
1. Replace MC PCBA.
2. Replace PM PCBA.
Backup audible
alarm fails to
sound for at least
two minutes
1. Replace MC PCBA.
2. Replace CPU PCBA.
Power switch
turned off but
backup alarm
remains on
1. Replace power switch overlay.
2. Replace MC PCBA.