Chapter 8
Performance Verification
1049766 Rev A
V60 Ventilator Service Manual
8-33
8.5.9
Alarms (Test 9)
This test verifies the performance of the alarm outputs.
Required equipment:
•
22-mm connector
•
Pressure pick-off port
•
18-in. smooth-bore tubing
•
Proximal pressure line assembly
•
Whisper Swivel
•
Ingmar test lung
•
Certifier FA Plus analyzer
•
DMM
•
Remote alarm test cable
•
Remote alarm test cable adapter
1. Connect a patient circuit, including test lung, Whisper Swivel, and
analyzer to the ventilator (Figure 8-22).
Figure 8-22: Alarms Test Configuration
2. Set up the test lung with an Rp20 and a compliance of 20.
3. If the ventilator is not already in normal ventilation mode, cycle power
to the ventilator to enter normal ventilation mode.
Содержание Respironics V60