Chapter 8
Performance Verification
1049766 Rev A
V60 Ventilator Service Manual
8-11
2. Use the up and down arrows to set these values:
Number of Breaths Averaged:
5
.
Second Average for Real-Time Values:
2.0
.
3. Touch
OK
when complete.
8.4.3
Trigger Options
Use Trigger Options to define how the analyzer detects the start of the
inspiratory and expiratory breath cycles.
1. Touch the Trigger options button to display the Trigger Options menu
(Figure 8-4).
Figure 8-4: Trigger Options Menu (Certifier FA Plus)
2. Select
TTL
.
3. Use the arrow buttons to select
Start
Trigger
.
4. Touch
OK
when complete.
Trigger options button
Содержание Respironics V60
Страница 1: ...Respironics V60 Ventilator Service Manual ...
Страница 8: ...viii V60 Ventilator Service Manual 1049766 Rev A Table of Contents This page is intentionally blank ...
Страница 204: ...Chapter 8 8 50 V60 Ventilator Service Manual 1049766 Rev A Performance Verification This page is intentionally blank ...
Страница 220: ...Chapter 9 9 16 V60 Ventilator Service Manual 1049766 Rev A Replacement Parts List This page is intentionally blank ...
Страница 242: ...Appendix A A 14 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...
Страница 244: ...Appendix B B 2 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...