Chapter 8
Performance Verification
1049766 Rev A
V60 Ventilator Service Manual
8-19
System Leak Test
Required equipment:
•
9
/
16
-
3
/
4
in. tapered plug (2)
•
Clamp (hemostat)
•
Pressure pick-off port
•
System leak test syringe
1. Remove the right side cover, air inlet filter retainer, and air inlet
housing to access the GDS (section 7.20).
2. Plug the air inlet, gas outlet, and patient pressure ports. Verify that the
system leak test syringe is pulled back, then connect it to the internal
exhalation port.
3. If the ventilator is not already in Diagnostic mode, press and hold the
Enter
button on the nav-ring and turn the ventilator on by pressing the
On/Shutdown
button on the UI. Within 5 seconds release and press
the nav-ring
Enter
button again to enter the Diagnostics menu.
4. Touch
Service
,
Misc
, then touch
System Leak Test
and follow the
onscreen prompts (Figure 8-13).
5. Touch
Close
, then remove the plugs and system leak test syringe.
6. Reinstall the inlet filter housing, inlet filter retainer, and the right side
cover (Chapter 7).
7. The system leak test is complete.
Figure 8-13: System Leak Test Screen
Содержание Respironics V60
Страница 1: ...Respironics V60 Ventilator Service Manual ...
Страница 8: ...viii V60 Ventilator Service Manual 1049766 Rev A Table of Contents This page is intentionally blank ...
Страница 204: ...Chapter 8 8 50 V60 Ventilator Service Manual 1049766 Rev A Performance Verification This page is intentionally blank ...
Страница 220: ...Chapter 9 9 16 V60 Ventilator Service Manual 1049766 Rev A Replacement Parts List This page is intentionally blank ...
Страница 242: ...Appendix A A 14 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...
Страница 244: ...Appendix B B 2 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...