Chapter 8
Performance Verification
1049766 Rev A
V60 Ventilator Service Manual
8-47
Pressure at 60 cmH
2
O
Analyzer pressure display (54 - 66 cmH
2
O)
cmH
2
O
cmH
2
O
cmH
2
O
Avg
Machine
display (54 - 66 cmH
2
O)
cmH
2
O
cmH
2
O
cmH
2
O
Avg
Proximal
display (54 - 66 cmH
2
O)
cmH
2
O
cmH
2
O
cmH
2
O
Test 5: Air Delivery/Flow Accuracy
Pass value
Fail value
Value after
repair
Air Delivery
Analyzer pressure display (> 64.42 cmH
2
O)
cmH
2
O
cmH
2
O
cmH
2
O
Analyzer flow display (> 136.8 SLPM)
SLPM
SLPM
SLPM
Air flow at 10 SLPM
Analyzer flow display (9 - 11 SLPM)
SLPM
SLPM
SLPM
Average
Air
display (9 - 11 SLPM)
SLPM
SLPM
SLPM
Air flow at 120 SLPM
Analyzer flow display (108 - 132 SLPM)
SLPM
SLPM
SLPM
Avg
Air
display (108 -132 SLPM)
SLPM
SLPM
SLPM
Air flow at 230 SLPM (N/A if above 6400 feet)
Analyzer flow display (207 - 253 SLPM)
SLPM
SLPM
SLPM
Avg
Air
display (207 - 253 SLPM)
SLPM
SLPM
SLPM
Analyzer flow display drops more than 15 SLPM
PASS
FAIL
Test 6: Oxygen Flow Accuracy
Pass value
Fail value
Value after
repair
Oxygen flow at 10 SLPM
Analyzer flow display (9 - 11 SLPM)
SLPM
SLPM
SLPM
Avg
O2
display (9 - 11 SLPM)
SLPM
SLPM
SLPM
Oxygen flow at 140 SLPM
Analyzer flow display (126 - 154 SLPM)
SLPM
SLPM
SLPM
Avg
O2
display (126 - 154 SLPM)
SLPM
SLPM
SLPM
Test 4: Pressure Accuracy
Pass value
Fail value
Value after
repair
Содержание Respironics V60
Страница 1: ...Respironics V60 Ventilator Service Manual ...
Страница 8: ...viii V60 Ventilator Service Manual 1049766 Rev A Table of Contents This page is intentionally blank ...
Страница 204: ...Chapter 8 8 50 V60 Ventilator Service Manual 1049766 Rev A Performance Verification This page is intentionally blank ...
Страница 220: ...Chapter 9 9 16 V60 Ventilator Service Manual 1049766 Rev A Replacement Parts List This page is intentionally blank ...
Страница 242: ...Appendix A A 14 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...
Страница 244: ...Appendix B B 2 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...