Chapter 8
8-8
V60 Ventilator Service Manual
1049766 Rev A
Performance Verification
8.3 View and Record
Ventilator Information
1. If the ventilator is not already in diagnostic mode, press and hold the
nav-ring
Enter
button, and press the
On/Shutdown
button on the user
interface. Within 5 seconds of power up, release and press the nav-
ring
Enter
button again to enter the Diagnostics menu.
2. Touch
Service
to display the Ventilator Information screen (Figure 8-
1), and record the following information on the Electrical Safety
(section 8.8,
Electrical Safety Data Form
) or Performance Verification
Data Form (section 8.9,
Performance Verification Data Form
):
Figure 8-1: Ventilator Information Screen
3. Touch
Pneumatics
and record the displayed barometric pressure.
4. Measure and record the ambient temperature and relative humidity.
5. Complete a diagnostic report (DRPTA) download (Chapter 6).
6. Turn the ventilator OFF.
•
Software options
•
Ventilator serial number
•
Software version
•
Total power-on hours
•
Hours since last PM
•
Battery lot number
•
Battery manufacture date
Содержание Respironics V60
Страница 1: ...Respironics V60 Ventilator Service Manual ...
Страница 8: ...viii V60 Ventilator Service Manual 1049766 Rev A Table of Contents This page is intentionally blank ...
Страница 204: ...Chapter 8 8 50 V60 Ventilator Service Manual 1049766 Rev A Performance Verification This page is intentionally blank ...
Страница 220: ...Chapter 9 9 16 V60 Ventilator Service Manual 1049766 Rev A Replacement Parts List This page is intentionally blank ...
Страница 242: ...Appendix A A 14 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...
Страница 244: ...Appendix B B 2 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...