Chapter 10
Specifications
1049766 Rev A
V60 Ventilator Service Manual
10-3
10.2 Patient data
Table 10-3: Patient data ranges, resolutions, and accuracies
Parameter
Range
Resolution
Accuracy
Patient data window
Breath phase/trigger
indicator
Spont, Timed,
Exhale
Color-coded display: Spont - turquoise,
Timed - orange, Exhale - blue
N/A
PIP
0 to 50 cmH
2
O
1 cmH
2
O
± 2 cmH
2
O
Pt. Leak
0 to 200 L/min
BTPS
1 L/min
N/A
Pt. Trig
0 to 100%
1%
± 10%
Rate
0 to 90 BPM
1 BPM
± 1 BPM
T
I
/T
TOT
0% to 91%
1%
± 5%
Tot.Leak
0 to 200 L/min
BTPS
1 L/min
N/A
V
E
.
0 to 99.0 L/min
BTPS
0.1 L/min
± 15% or 0.3 L/min (whichever is great-
er)
V
T
0 to 3000 mL
BTPS
5 mL
± 15% for volumes above 200 mL
Waveform window
P
waveform
0 to 50 cmH
2
O
Time axis: 1 second
N/A
V
E
waveform
.
-240 to 240 L/
min BTPS
Time axis: 1 second
N/A
V
waveform
0 to 3000 mL
BTPS
Time axis: 1 second
N/A
Содержание Respironics V60
Страница 1: ...Respironics V60 Ventilator Service Manual ...
Страница 8: ...viii V60 Ventilator Service Manual 1049766 Rev A Table of Contents This page is intentionally blank ...
Страница 204: ...Chapter 8 8 50 V60 Ventilator Service Manual 1049766 Rev A Performance Verification This page is intentionally blank ...
Страница 220: ...Chapter 9 9 16 V60 Ventilator Service Manual 1049766 Rev A Replacement Parts List This page is intentionally blank ...
Страница 242: ...Appendix A A 14 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...
Страница 244: ...Appendix B B 2 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...