Chapter 8
Performance Verification
1049766 Rev A
V60 Ventilator Service Manual
8-49
Test 11: Internal Battery
Circle one
Transition from AC to battery operation does not interrupt
ventilator operation and indicators operate correctly
PASS
FAIL
N/A
Ventilator powers up in normal ventilation and indicators
operate correctly
PASS
FAIL
N/A
Transition from battery to AC operation does not interrupt
ventilator operation and indicators operate correctly
PASS
FAIL
N/A
Battery indicator stops flashing
PASS
FAIL
N/A
Internal battery fully charged
PASS
FAIL
N/A
(Internal Battery is not fully charged if Battery Low message is displayed during normal
operation on AC power: performance verification result is
LIMITED USE
)
Did the performance verification pass? (circle one)
YES
NO
LIMITED USE*
*
LIMITED USE
indicates that the internal battery operational check failed: recommend that
the operator connect the ventilator to AC power for at least 5 hours, then verify battery
performance.
Technician’s signature
Date
Содержание Respironics V60
Страница 1: ...Respironics V60 Ventilator Service Manual ...
Страница 8: ...viii V60 Ventilator Service Manual 1049766 Rev A Table of Contents This page is intentionally blank ...
Страница 204: ...Chapter 8 8 50 V60 Ventilator Service Manual 1049766 Rev A Performance Verification This page is intentionally blank ...
Страница 220: ...Chapter 9 9 16 V60 Ventilator Service Manual 1049766 Rev A Replacement Parts List This page is intentionally blank ...
Страница 242: ...Appendix A A 14 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...
Страница 244: ...Appendix B B 2 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...