Chapter 7
7-28
V60 Ventilator Service Manual
1049766 Rev A
Component Removal/Installation
7.25 Oxygen Solenoid
Valve
Follow these steps to remove the oxygen solenoid valve (Figure 7-26). Reverse
to install.
1. Remove the GDS (section 7.20).
2. Disconnect the oxygen solenoid cable from the DA PCBA.
3. Remove the screws that attach the solenoid orientation clip to the
solenoid mount.
NOTE:
When installing the oxygen solenoid, apply a thin coat of Krytox on
the o-rings.
Figure 7-26: Removing the Oxygen Solenoid Valve
Oxygen
solenoid
valve
Oxygen solenoid cable
Oxygen solenoid valve
DA PCBA
Solenoid
orientation
clip
Solenoid
mount
M3 x 8 screw (x3)
(62 + 5.7 N cm/
5.5 + 0.5 in.-lbf)
M2.5 x 6 screw (x2)
(34 + 3.4 N cm/
3.0 + 0.3 in.-lbf)
O-rings