Chapter 7
Component Removal/Installation
1049766 Rev A
V60 Ventilator Service Manual
7-17
7.16 Power Supply
Follow these steps to remove the power supply (Figure 7-16). Reverse to
install.
NOTE:
Do not remove any connections from the power supply terminal strip.
1. Remove MC PCBA (section 7.8) and PM PCBA (section 7.15).
2. Remove EMI shroud from power supply by slightly lifting the shroud
away from the power supply with a small screwdriver.
3. Remove the screws that attach the mounting plate to the base. (Power
supply includes attached mounting plate.)
4. Disconnect AC inlet cable and power harness.
NOTE:
Squeeze EMI shroud before reinstalling to ensure that it fits snugly over
the power supply.
Figure 7-16: Removing the Power Supply
Power supply
EMI shroud
Mounting plate
M3 x 6 screw (x4)
(62 + 5.7 N cm/
5.5 + 0.5 in.-lbf)
From under mounting plate (not shown):
M3 x 6 screw (x4)
(62 + 5.7 N cm/
5.5 + 0.5 in.-lbf)
Holes in EMI shroud fit over screw
heads on the side of the power supply
Содержание Respironics V60
Страница 1: ...Respironics V60 Ventilator Service Manual ...
Страница 8: ...viii V60 Ventilator Service Manual 1049766 Rev A Table of Contents This page is intentionally blank ...
Страница 204: ...Chapter 8 8 50 V60 Ventilator Service Manual 1049766 Rev A Performance Verification This page is intentionally blank ...
Страница 220: ...Chapter 9 9 16 V60 Ventilator Service Manual 1049766 Rev A Replacement Parts List This page is intentionally blank ...
Страница 242: ...Appendix A A 14 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...
Страница 244: ...Appendix B B 2 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...