Chapter 7. Component Removal/Installation
This chapter describes how to remove and install the replaceable components
in the V60 ventilator.
WARNING:
To avoid personal injury, disconnect all power from the ventilator before
servicing or cleaning.
CAUTION:
Component removal and installation is to be performed only by a
qualified service technician.
CAUTION:
To avoid equipment damage due to static electricity, perform all repairs
in an antistatic, electrostatic discharge (ESD)-protected environment.
CAUTION:
Replacing the CPU PCBA deletes the serial number, significant event
log entries, and ventilator preferences. Review the significant event log
before removing the CPU PCBA.
CAUTION:
The gas delivery subsystem (GDS) must be leak tested if you replace any
of these components:
•
Data acquisition (DA) PCBA
•
Oxygen inlet filter
•
Oxygen solenoid valve
•
Oxygen filter cap