Chapter 5
5-18
V60 Ventilator Service Manual
1049766 Rev A
Diagnostic Mode and Troubleshooting
5.3 Touch Screen
Calibration
The Touch Screen Calibration button allows you to calibrate the touch screen x
and y coordinates (Figure 5-17):
1. From the Diagnostics Menu, touch
Touch
Screen
Calibration
.
NOTE:
If the
Touch Screen Calibration
button does not respond, press the nav-
ring
Enter
button to begin.
2. Follow the onscreen instructions to perform the calibration. Touch the
middle of each target with a blunt, narrow object.
Figure 5-17: Calibrate Touch Screen