Appendix A
1049766 Rev A
V60 Ventilator Service Manual
A-13
17. Remove the null modem cable and adapter from the V60 ventilator.
18. Turn the V60 ventilator off by pressing and holding the
On/Shutdown
button on the user interface.
19. Enter Diagnostic mode by pressing and holding the nav-ring
Enter
button and turn on the ventilator by pressing the
On/Shutdown
button
on the user interface. Within 5 seconds, release and press the nav-ring
Enter
button again to enter Diagnostic mode.
20. Touch
Service
and verify the Options on the V60 ventilator are the
same as the options listed on the GatewayInterface screen.
21. Click
Exit
on the GatewayInterface screen.
22. Turn the V60 ventilator off by pressing the
On/Shutdown
button.
23. Software reinitiation is complete.
A.4 Remote
Troubleshooting
Follow these steps to connect the V60 ventilator to Respi-Link and prepare for
remote troubleshooting.
NOTE:
Do not connect the null modem cable until instructed to do so.
1. Turn on the PC and verify the Gateway Interface is running by double-
clicking on
Start Auto Respi-Link Gateway
in the Respi-Link folder on
your desktop.
2. If the ventilator is not already in Diagnostic mode, press and hold the
nav-ring
Enter
button and turn on the ventilator by pressing the
On/Shutdown
button on the user interface. Within 5 seconds, press
the nav-ring
Enter
button again to enter Diagnostic mode.
3. Touch
Service
, then
Pneumatics
.
4. Connect a standard 9-pin male-female RS-232 null modem cable and
9-pin to 25-pin female-male adapter between the PC serial port and
the V60 ventilator.
5. The V60 ventilator is now connected to Respi-Link and can be
monitored remotely by a Respironics representative.
6. Contact your local service representative or Product Support at 800-
345-6443 for further instructions.
Содержание Respironics V60
Страница 1: ...Respironics V60 Ventilator Service Manual ...
Страница 8: ...viii V60 Ventilator Service Manual 1049766 Rev A Table of Contents This page is intentionally blank ...
Страница 204: ...Chapter 8 8 50 V60 Ventilator Service Manual 1049766 Rev A Performance Verification This page is intentionally blank ...
Страница 220: ...Chapter 9 9 16 V60 Ventilator Service Manual 1049766 Rev A Replacement Parts List This page is intentionally blank ...
Страница 242: ...Appendix A A 14 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...
Страница 244: ...Appendix B B 2 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...