Chapter 8
Performance Verification
1049766 Rev A
V60 Ventilator Service Manual
8-5
8.1 Required Test
Equipment
Table 8-3 summarizes the test equipment required, and Table 8-4 summarizes
the service accessories required for performance verification. Check the
calibration status of all test equipment before use.
Table 8-3: Required Test Equipment for Performance Verification
Description
Recommended Manufacturer/Model
Electrical safety analyzer
Dale LT 5440 or equivalent
Pneumatic calibration analyzer(s)
capable of measuring low pressure
(cmH
2
O), flow rate (LPM), volume
(liters), and respiratory rate
TSI Certifier FA Plus (P/N 1040311),
or equivalent
Oxygen analyzer
TSI Certifier FA Plus or equivalent
Pressure analyzer capable of
measuring high pressure (PSI)
TSI Certifier FA Plus or equivalent
Digital multimeter (DMM) and
frequency counter
Local supplier
Test lung
IngMar QuickLung or equivalent
Temperature/humidity monitor
Fisher Scientific 11-661-14 or
equivalent
NOTE:
An oxygen source capable of delivering 140 LPM (40-87 PSI) is
required for oxygen system tests.
Содержание Respironics V60
Страница 1: ...Respironics V60 Ventilator Service Manual ...
Страница 8: ...viii V60 Ventilator Service Manual 1049766 Rev A Table of Contents This page is intentionally blank ...
Страница 204: ...Chapter 8 8 50 V60 Ventilator Service Manual 1049766 Rev A Performance Verification This page is intentionally blank ...
Страница 220: ...Chapter 9 9 16 V60 Ventilator Service Manual 1049766 Rev A Replacement Parts List This page is intentionally blank ...
Страница 242: ...Appendix A A 14 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...
Страница 244: ...Appendix B B 2 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...