Chapter 6
Reports and Software Downloads
1049766 Rev A
V60 Ventilator Service Manual
6-17
6.2 Generating a
Diagnostic Report
(DRPTA)
The DRPTA command allows you to save ventilator diagnostic codes to your
PC.
NOTE:
Do not connect the null modem cable until instructed to do so.
1. Place the ventilator in diagnostic mode.
2. Connect the 25-pin to 9-pin adapter and a null modem cable between
the ventilator COM port and the service PC.
3. Click on the HyperTerminal or Tera Term icon to launch the program
that communicates with the V60 ventilator.
4. Type
DRPTA
(all caps) in the dialog box (Figure 6-31), and press
Enter
.
Figure 6-31: Entering the DRPTA Command
HyperTerminal dialog box
Tera Term dialog box