1049766 Rev A
V60 Ventilator Service Manual
5-1
Chapter 5. Diagnostic Mode and
Troubleshooting
Use diagnostic mode to perform these functions:
•
System settings
: language, ventilator date and time, pressure units,
restore default settings, enable software options, and Baud rate.
•
Service
: ventilator information, pneumatics and outputs screens (for
troubleshooting), and miscellaneous features (viewing the significant
event log, downloading software, touch screen diagnostics, high
pressure leak test, and system leak test).
•
Touch screen calibration
.
This chapter also provides detailed information on diagnostic codes and
troubleshooting (section 5.4).
WARNING:
Do not enter diagnostic mode while a patient is connected to the ventilator.
NOTE:
This mode is primarily reserved for use by a qualified service technician.
To enter diagnostic mode:
1. Press and hold the nav-ring
Enter
button then press and release the
ventilator
On/Shutdown
button on the user interface.
2. Within five seconds, press the nav-ring
Enter
button again to enter the
Diagnostics Menu (Figure 5-1).
Figure 5-1:
Diagnostics Menu
3. To exit diagnostic mode at any time, press the
ON/Shutdown
button.
Содержание Respironics V60
Страница 1: ...Respironics V60 Ventilator Service Manual ...
Страница 8: ...viii V60 Ventilator Service Manual 1049766 Rev A Table of Contents This page is intentionally blank ...
Страница 204: ...Chapter 8 8 50 V60 Ventilator Service Manual 1049766 Rev A Performance Verification This page is intentionally blank ...
Страница 220: ...Chapter 9 9 16 V60 Ventilator Service Manual 1049766 Rev A Replacement Parts List This page is intentionally blank ...
Страница 242: ...Appendix A A 14 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...
Страница 244: ...Appendix B B 2 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...