Chapter 5
Diagnostic Mode and Troubleshooting
1049766 Rev A
V60 Ventilator Service Manual
5-21
1107
(POST)
Proximal pressure sensor calibration data error.
Proximal pressure is not measured. Pressure-related alarms
are compromised.
1. Replace DA PCBA.
1108
(POST)
1109
(CBIT)
Machine pressure sensor auto-zero failed.
Proximal pressure is not measured. Pressure-related alarms
are compromised.
1. Verify pin alignment between
solenoids and DA PCBA.
2. Replace machine auto-zero
solenoid (SOL 2).
3. Replace DA PCBA.
110A
(POST)
110B
(CBIT)
Proximal pressure sensor auto-zero failed.
Proximal pressure is not measured. Pressure-related alarms
are compromised.
1. Verify pin alignment between
solenoids and DA PCBA.
2. Replace proximal auto-zero
solenoid (SOL 3).
3. Replace DA PCBA.
110C
(CBIT)
Machine pressure sensor range error.
Proximal pressure is not measured. Pressure-related alarms
are compromised.
1. Cycle ventilator power to reset
proximal pressure sensor.
2. Replace DA PCBA.
110D
(CBIT)
Proximal pressure sensor range error.
Proximal pressure is not measured. Pressure-related alarms
are compromised.
1. Cycle ventilator power to reset
proximal pressure sensor.
2. Replace DA PCBA.
110E
(POST)
Air flow sensor calibration data error.
Flow-related patient data is disabled. Oxygen concentration
switches to 21% (air only). Default volume used in AVAPS
mode. Standby disabled. Volume, leak, disconnect, and
occlusion alarms compromised.
1. Replace flow sensor cable.
2. Replace flow sensor assembly.
3. Replace DA PCBA.
110F
(POST)
Oxygen flow sensor calibration data error.
Ventilation continues with air only.
1. Replace flow sensor cable.
2. Replace flow sensor assembly.
3. Replace DA PCBA.
1110
(POST)
Oxygen pressure sensor calibration data error.
Ventilation continues with air only.
1. Replace DA PCBA.
Table 5-1: Diagnostic Codes and Troubleshooting
Diagnostic
code
Description
Recommended repair
Содержание Respironics V60
Страница 1: ...Respironics V60 Ventilator Service Manual ...
Страница 8: ...viii V60 Ventilator Service Manual 1049766 Rev A Table of Contents This page is intentionally blank ...
Страница 204: ...Chapter 8 8 50 V60 Ventilator Service Manual 1049766 Rev A Performance Verification This page is intentionally blank ...
Страница 220: ...Chapter 9 9 16 V60 Ventilator Service Manual 1049766 Rev A Replacement Parts List This page is intentionally blank ...
Страница 242: ...Appendix A A 14 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...
Страница 244: ...Appendix B B 2 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...