Chapter 4
4-6
V60
Ventilator
Service Manual
1049766 Rev A
Periodic Maintenance
19. Type
#CLRLOG
(all caps) in the dialog box, and press
Enter
to clear the diagnostic log
(Figure 4-5).
Figure 4-5: Clear Diagnostic Codes Command (HyperTerminal)
20. When the diagnostic log has been successfully cleared, the dialog box
displays:
?CLRLOGOK
21. Touch
System
Settings
, and then touch
Restore
Default
Settings
.
22. Touch
Date/Time
and correct the date and time if necessary.
23. Turn the V60 ventilator off by pressing the
ON/Shutdown
button.
24. Upon successful completion of the performance verification, hours
reset, and restoring the default settings, enter the required
information on the PM and electrical safety labels.