7.27 Blower
Follow these steps to remove the blower (Figure 7-28). Reverse to install.
1. Remove the top cover (section 7.4).
2. Disconnect the blower cable from the MC PCBA.
3. Remove the boot and hose clamps that attach the blower to the gas
outlet.
4. Remove the shoulder screws that attach the blower to the base.
5. Loosen the clamps on the boot that attaches the blower to the GDS,
then remove the blower, boot, and clamps.
NOTE:
When installing the blower:
•
Ensure that the boot is fully and clamped to the blower before
installing the blower.
•
Install new motor mount grommets when installing a new blower.
Figure 7-28: Removing the Blower
Hose clamps (x2)
Hose clamps
(x2)
Shoulder screw (x4)
(113 + 11.3 N cm/
10 + 1 in.-lbf)
Blower
Blower
Manifold to
blower boot
Manifold to
blower boot
Blower
cable to
MC PCBA
Gas outlet
To gas outlet
To GDS
Содержание Respironics V60
Страница 1: ...Respironics V60 Ventilator Service Manual ...
Страница 8: ...viii V60 Ventilator Service Manual 1049766 Rev A Table of Contents This page is intentionally blank ...
Страница 204: ...Chapter 8 8 50 V60 Ventilator Service Manual 1049766 Rev A Performance Verification This page is intentionally blank ...
Страница 220: ...Chapter 9 9 16 V60 Ventilator Service Manual 1049766 Rev A Replacement Parts List This page is intentionally blank ...
Страница 242: ...Appendix A A 14 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...
Страница 244: ...Appendix B B 2 V60 Ventilator Service Manual 1049766 Rev A This page is intentionally blank ...