Chapter 7
7-26
V60 Ventilator Service Manual
1049766 Rev A
Component Removal/Installation
Figure 7-24: Removing the DA PCBA
M3 x 8 screw (x7)
(62 + 5.7 N cm/
5.5 + 0.5 in.-lbf)
M3 washer (x7)
DA PCBA installed on GDS
Oxygen solenoid
cable
DA to MC ribbon
cable connector
Underside of installed DA PCBA:
solenoid pins (do not bend)
Underside of DA PCBA: ensure that transducer grommets
and o-rings are installed at reassembly
O-ring (x2)
Flow sensor ribbon
cable assembly
Transducer grommet (x2)
This grommet
is
correctly
installed
This grommet is
NOT
correctly installed