Cypher SPM
User Guide
8. Select the AC mode cantilever holder.
l
Identify the correct cantilever holder.
l
This demo uses the standard AC Air holder, part#
901.705.
Tip:
To learn more about cantilever holders for the
standard scanner, see
Visual identification of SS can-
9. Prepare the cantilever mounting workspace.
l
Set out your changing station, tweezers, and can-
tilevers on a clean, well lighted surface.
l
Make sure that the changing station is labeled “Air”
(there is also a “Droplet” changing station for the
droplet holder.)
l
A low power binocular dissection stereoscope with
light source can be useful for some of the following
steps.
Tip:
Cleaning the tweezer tips with alcohol
improves the handling of the cantilevers.
10. Mount the cantilever holder in the changing station.
l
Carefully insert the cantilever holder as shown.
l
The V-shaped piece of metal on the back of the
holder slides into the dovetail joint on the changing
station. The cantilever should be pointing down.
l
If the cantilever holder does not slide in easily,
loosen the screw on the clamping mechanism.
30 Workflows
Page 201 of 243
Содержание Cypher SPM
Страница 1: ...Cypher SPM User Guide Part No XX XXXX XXX...
Страница 8: ...Part I Getting Started System safety and overview...
Страница 165: ...Cypher SPM User Guide Sub Figure a First view of new cantilever 24 Engaging the cantilever on the sample Page 165 of 243...
Страница 167: ...Cypher SPM User Guide Sub Figure c Cantilever in focus See 9 24 Engaging the cantilever on the sample Page 167 of 243...
Страница 169: ...Cypher SPM User Guide Sub Figure a Cantilever Zoom See 10 24 Engaging the cantilever on the sample Page 169 of 243...
Страница 170: ...Cypher SPM User Guide Sub Figure b Laser Spot ON See 11 Page 170 of 243 24 Engaging the cantilever on the sample...
Страница 173: ...Cypher SPM User Guide Sub Figure a Offset 24 Engaging the cantilever on the sample Page 173 of 243...
Страница 176: ...Cypher SPM User Guide Sub Figure a Master Panel Main Tab Page 176 of 243 24 Engaging the cantilever on the sample...
Страница 186: ...Cypher SPM User Guide Sub Figure a Before the Move Page 186 of 243 25 AC mode imaging...
Страница 189: ...Cypher SPM User Guide 26 26 Contact Mode Imaging In Air 26 Contact mode imaging in air Page 189 of 243...
Страница 216: ...Part VI Shipping and Maintenance This part of the user guide covers moving and shipping the instrument...