Prepare to land the tip:
l
Click the Move to Pre-Engage button. Motors automatically bring the tip to 50μm from the sur-
face.
WARNING:
If you set a bad sample height and/or tip position you may ram your cantilever into the sample
and break it. A firmware safety feature immediately cuts motor power when the optical detector
fails to measure reflected light from the broken lever. This prevents the cantilever holder from
ramming the sample.
18.
Fine tune spot position:
l
At this point you should adjust the spot position so that it is near the end of the cantilever using
the controls in the upper left hand corner of the video panel. Note that a decrease in Sum Signal
indicates that light is spilling off the cantilever. The latter is undesirable and should be avoided.
HINT
For fine positioning hold down the SHIFT key on your keyboard while clicking the onscreen but-
tons which move the cantilever under the laser spot (See
).
19.
Center the laser on the photo detector:
l
Hit the red button at the center of the video window motion controls and watch the Deflection
Signal go to zero. This action causes small motors to steer the laser beam to the optimum pos-
ition on Cypher's photo detector.
Page 164 of 243
24 Engaging the cantilever on the sample
Содержание Cypher SPM
Страница 1: ...Cypher SPM User Guide Part No XX XXXX XXX...
Страница 8: ...Part I Getting Started System safety and overview...
Страница 165: ...Cypher SPM User Guide Sub Figure a First view of new cantilever 24 Engaging the cantilever on the sample Page 165 of 243...
Страница 167: ...Cypher SPM User Guide Sub Figure c Cantilever in focus See 9 24 Engaging the cantilever on the sample Page 167 of 243...
Страница 169: ...Cypher SPM User Guide Sub Figure a Cantilever Zoom See 10 24 Engaging the cantilever on the sample Page 169 of 243...
Страница 170: ...Cypher SPM User Guide Sub Figure b Laser Spot ON See 11 Page 170 of 243 24 Engaging the cantilever on the sample...
Страница 173: ...Cypher SPM User Guide Sub Figure a Offset 24 Engaging the cantilever on the sample Page 173 of 243...
Страница 176: ...Cypher SPM User Guide Sub Figure a Master Panel Main Tab Page 176 of 243 24 Engaging the cantilever on the sample...
Страница 186: ...Cypher SPM User Guide Sub Figure a Before the Move Page 186 of 243 25 AC mode imaging...
Страница 189: ...Cypher SPM User Guide 26 26 Contact Mode Imaging In Air 26 Contact mode imaging in air Page 189 of 243...
Страница 216: ...Part VI Shipping and Maintenance This part of the user guide covers moving and shipping the instrument...