Cypher SPM
User Guide
27.2.1 Install the sample on the scanner and connect the bias lead
1. Place the sample on the scanner stage and connect the bias lead.
l
Position the sample so that the magnet is on the
right hand side of the scanner to prevent inter-
ference with the cantilever holder.
l
Attach one end of the bias wire to the magnetic con-
tact on the scanner top.
l
Place the other end of the lead on the magnet. The
lead is magnetic and will stick to the magnet when
it's close enough.
Note:
The scanner cap is hard anodized aluminum
and will insulate the the sample puck so bias
voltages up to +/-10v can be directly connected
without additional insulation.
2. Adding resistance in the bias voltage path.
l
In cases where your sample is highly conductive, you
may want to add a known resistance to keep the OR
CA amplifier from saturating. An example of this is
the HOPG sample provided in the kit.
l
Substitute the Bias lead with a bias lead including a
resistor.
Note:
The ORCA holder kit includes a 500 MΩ test
resistor. If you are planning on scanning the HOPG
sample to practice using the ORCA holder, use the
test resistor instead of the bias lead.
Plug one end of a bias voltage lead
into the sample socket on the ter-
minal block
.
27.3 Mounting the cantilever
Mounting a cantilever is the same procedure as is used in all other AFM applications with the excep-
tion of lever type. Conducting AFM (ORCA) requires a conductive path between the tip and the can-
tilever spring clip. The ORCA kit includes a sample pack of 10 Electrilevers. Additional levers can be
purchased from Asylum Research.
For details of mounting the cantilever, see
Loading the cantilever and sample on a standard scan-
Page 192 of 243
27 Preparing for conductive AFM imaging on the standard scanner
Содержание Cypher SPM
Страница 1: ...Cypher SPM User Guide Part No XX XXXX XXX...
Страница 8: ...Part I Getting Started System safety and overview...
Страница 165: ...Cypher SPM User Guide Sub Figure a First view of new cantilever 24 Engaging the cantilever on the sample Page 165 of 243...
Страница 167: ...Cypher SPM User Guide Sub Figure c Cantilever in focus See 9 24 Engaging the cantilever on the sample Page 167 of 243...
Страница 169: ...Cypher SPM User Guide Sub Figure a Cantilever Zoom See 10 24 Engaging the cantilever on the sample Page 169 of 243...
Страница 170: ...Cypher SPM User Guide Sub Figure b Laser Spot ON See 11 Page 170 of 243 24 Engaging the cantilever on the sample...
Страница 173: ...Cypher SPM User Guide Sub Figure a Offset 24 Engaging the cantilever on the sample Page 173 of 243...
Страница 176: ...Cypher SPM User Guide Sub Figure a Master Panel Main Tab Page 176 of 243 24 Engaging the cantilever on the sample...
Страница 186: ...Cypher SPM User Guide Sub Figure a Before the Move Page 186 of 243 25 AC mode imaging...
Страница 189: ...Cypher SPM User Guide 26 26 Contact Mode Imaging In Air 26 Contact mode imaging in air Page 189 of 243...
Страница 216: ...Part VI Shipping and Maintenance This part of the user guide covers moving and shipping the instrument...