Cypher SPM
User Guide
2. Be careful not to poke the sample stage diaphragms.
l
Be careful never to poke the sample stage dia-
phragms with tweezers or screwdrivers as this may
cause a hole in the diaphragm.
l
Some versions of the sample stages have per-
manently attached diaphragms. These stages cannot
be repaired if their diaphragms are damaged.
3. Be careful when reattaching the cantilever clip not to crack the glass.
l
Anytime the cantilever clip has been completely
removed - for cleaning - care must be taken when
reattaching the clip.
l
Do not tighten the screw until you are sure that the
clip is aligned correctly and has dropped into the
pocket in the glass body.
Failure to align the cantilever clip with the pocket
before tightening the clip will break the cantilever
holder glass.
4. Use extra care when loading the Heater sample stage.
l
The heater sample stage is extremely fragile. Only
use minimal force when seating the sample stage
dovetail against the scanner.
If you apply too much force to the heater stage
you will break it.
l
If you use your finger to push the stage down into
the scanner, you should only need enough pressure
to extend the diaphragm. Any more pressure will
break the heater. If it is taking more pressure to push
the dovetail into the scanner, simply push the dove-
tail over (with a wrench from the bottom of the
sample stage) until it drops into the scanner.
l
It is a good idea to wear gloves when doing this so
that you minimize finger grease getting cooked onto
the heater stage.
10 Cell Body and Sample Stage Guide
Page 78 of 243
Содержание Cypher SPM
Страница 1: ...Cypher SPM User Guide Part No XX XXXX XXX...
Страница 8: ...Part I Getting Started System safety and overview...
Страница 165: ...Cypher SPM User Guide Sub Figure a First view of new cantilever 24 Engaging the cantilever on the sample Page 165 of 243...
Страница 167: ...Cypher SPM User Guide Sub Figure c Cantilever in focus See 9 24 Engaging the cantilever on the sample Page 167 of 243...
Страница 169: ...Cypher SPM User Guide Sub Figure a Cantilever Zoom See 10 24 Engaging the cantilever on the sample Page 169 of 243...
Страница 170: ...Cypher SPM User Guide Sub Figure b Laser Spot ON See 11 Page 170 of 243 24 Engaging the cantilever on the sample...
Страница 173: ...Cypher SPM User Guide Sub Figure a Offset 24 Engaging the cantilever on the sample Page 173 of 243...
Страница 176: ...Cypher SPM User Guide Sub Figure a Master Panel Main Tab Page 176 of 243 24 Engaging the cantilever on the sample...
Страница 186: ...Cypher SPM User Guide Sub Figure a Before the Move Page 186 of 243 25 AC mode imaging...
Страница 189: ...Cypher SPM User Guide 26 26 Contact Mode Imaging In Air 26 Contact mode imaging in air Page 189 of 243...
Страница 216: ...Part VI Shipping and Maintenance This part of the user guide covers moving and shipping the instrument...