Starting And Stopping The System
5.1 System power up
Note:
Before you start, familiarize yourself with the safety section, see
1. Power up.
l
Boot up the computer.
l
Turn on the Cypher SPM by depressing the power
switch on the front of the ARC2 controller.
If everything is working correctly, two different green
LEDs are illuminated. The first LED is located above the
power switch on the ARC2 and the second LED (labeled
“power”) is located at the front of the Cypher enclos-
ure.
2. Check that the laser key on the ARC2 controller is in the ON position.
The red LED labeled “laser” at the front of the Cypher enclosure should be illuminated and the
software initializes.
Make sure your fingers are clear from any pinch points before homing the motors, see
Avoiding unsafe situations on page 11
. You can abort the homing process at any time by press-
ing and holding down the Esc key located at the top left corner of the keyboard. If you abort
the homing process before it is finished, you will not be able operate the motors.
5.2 System power down
The following procedure should be used if the Cypher will not be used for some time, for
instance, at the end of the workday.
1. Once you are done imaging, save your data to a desired directory. Close Igor and shut
down the computer.
2. The tip will disengage automatically when imaging stops, but for added safety, motor the
tip away from the sample. You may want to remove the sample at this point.
5 Starting and stopping the system
Page 30 of 243
Содержание Cypher SPM
Страница 1: ...Cypher SPM User Guide Part No XX XXXX XXX...
Страница 8: ...Part I Getting Started System safety and overview...
Страница 165: ...Cypher SPM User Guide Sub Figure a First view of new cantilever 24 Engaging the cantilever on the sample Page 165 of 243...
Страница 167: ...Cypher SPM User Guide Sub Figure c Cantilever in focus See 9 24 Engaging the cantilever on the sample Page 167 of 243...
Страница 169: ...Cypher SPM User Guide Sub Figure a Cantilever Zoom See 10 24 Engaging the cantilever on the sample Page 169 of 243...
Страница 170: ...Cypher SPM User Guide Sub Figure b Laser Spot ON See 11 Page 170 of 243 24 Engaging the cantilever on the sample...
Страница 173: ...Cypher SPM User Guide Sub Figure a Offset 24 Engaging the cantilever on the sample Page 173 of 243...
Страница 176: ...Cypher SPM User Guide Sub Figure a Master Panel Main Tab Page 176 of 243 24 Engaging the cantilever on the sample...
Страница 186: ...Cypher SPM User Guide Sub Figure a Before the Move Page 186 of 243 25 AC mode imaging...
Страница 189: ...Cypher SPM User Guide 26 26 Contact Mode Imaging In Air 26 Contact mode imaging in air Page 189 of 243...
Страница 216: ...Part VI Shipping and Maintenance This part of the user guide covers moving and shipping the instrument...