Use only the original cartons and foam for shipping this delicate equipment. If you do not
have the cartons described in this chapter, contact Asylum Research and request empty cartons
with foam.
31.3 Instrument weight
The Cypher instrument is made of many heavy metal parts. Be prepared to have two people on
hand whenever lifting of the empty enclosure is required.
Note:
Even the scanner module is quite heavy. Be prepared to support its weight as you pull it
from the support rails.
31.4 Complete system shutdown
1. Motor up the cantilever holder.
2. Remove any samples from the scanner.
3. Remove all loose items from the AFM enclosure.
4. Shut down the IGOR software and the computer.
5. Power off the AFM controller.
6. Unplug all cables leading to the AFM enclosure and place them aside, keeping them together.
Never lift the AFM enclosure with the instrument inside. Don't try to move the AFM from the
lab to, say, the loading dock prior to packaging. Disassemble the AFM within its normal laboratory
setting.
31.5 Scanner removal and packaging
Call or otherwise contact Asylum Research and talk to us before moving anything. Chances are
that with a few phone calls and the help of this chapter you can do the job yourself. Not calling us
will likely result in instrument damage, costly repairs and downtime, and the voiding of your war-
ranty.
Materials
Part# 991.001: Carton 10" x 10" x 10"
Part# 990.035: Foam (scanner)
Tools
Part# 290.111: 0.050” Hex driver
Final weight
14 lbs / 6 kg
Minimum preparation System shutdown (see
Complete system shutdown above
).
31 Shipping or moving
Page 218 of 243
Содержание Cypher SPM
Страница 1: ...Cypher SPM User Guide Part No XX XXXX XXX...
Страница 8: ...Part I Getting Started System safety and overview...
Страница 165: ...Cypher SPM User Guide Sub Figure a First view of new cantilever 24 Engaging the cantilever on the sample Page 165 of 243...
Страница 167: ...Cypher SPM User Guide Sub Figure c Cantilever in focus See 9 24 Engaging the cantilever on the sample Page 167 of 243...
Страница 169: ...Cypher SPM User Guide Sub Figure a Cantilever Zoom See 10 24 Engaging the cantilever on the sample Page 169 of 243...
Страница 170: ...Cypher SPM User Guide Sub Figure b Laser Spot ON See 11 Page 170 of 243 24 Engaging the cantilever on the sample...
Страница 173: ...Cypher SPM User Guide Sub Figure a Offset 24 Engaging the cantilever on the sample Page 173 of 243...
Страница 176: ...Cypher SPM User Guide Sub Figure a Master Panel Main Tab Page 176 of 243 24 Engaging the cantilever on the sample...
Страница 186: ...Cypher SPM User Guide Sub Figure a Before the Move Page 186 of 243 25 AC mode imaging...
Страница 189: ...Cypher SPM User Guide 26 26 Contact Mode Imaging In Air 26 Contact mode imaging in air Page 189 of 243...
Страница 216: ...Part VI Shipping and Maintenance This part of the user guide covers moving and shipping the instrument...