There are a variety of ORCA cantilever holders each based on either a single or dual amplification
design. The design type and amplification gain are labeled on the top of the holder. Like all the
Cypher cantilever holders, a built-in circuit in the holder allows the software to automatically sense
the type of holder and configure the system accordingly.
The amplification range of the ORCA amplifier is expressed by its sensitivity—the ability to produce
a voltage output from a certain current flow into the tip. In terms of the full range of the ORCA amp-
lifier, the output is +/- 10 V, so multiplying the sensitivity by +/- 10 will give you the full range.
The ORCA amplifier incorporates the use of a trans-impedance amplifier which converts the input
current from the tip to an output voltage. The input potential of the amp is referenced to ground
so the tip is essentially held at 0 V potential. During the measurement, the sample can be biased
b/- 10 V using a voltage source provided by the Cypher electronics.
Each ORCA cantilever holder has a fixed gain(s) to provide the highest current measurement range
while considering the lowest noise. The following ORCA holders are currently available. Custom
holders can be configured on request.
Part number
Sensitivity Current Range Typical noise 1-1 KHz
901.730
2 nA/V
+/- 20 nA
1.5 pA
901.737
0.2 nA/V
+/- 2 nA
750 fA
901.708
1 nA/V
+/- 10 nA
3 pA
1 μA/V
+/- 10 μA
75 pA
7.5.1 Single gain
Here is a conceptual block diagram of the single gain ORCA amplifier.
7 Standard scanner cantilever holders
Page 40 of 243
Содержание Cypher SPM
Страница 1: ...Cypher SPM User Guide Part No XX XXXX XXX...
Страница 8: ...Part I Getting Started System safety and overview...
Страница 165: ...Cypher SPM User Guide Sub Figure a First view of new cantilever 24 Engaging the cantilever on the sample Page 165 of 243...
Страница 167: ...Cypher SPM User Guide Sub Figure c Cantilever in focus See 9 24 Engaging the cantilever on the sample Page 167 of 243...
Страница 169: ...Cypher SPM User Guide Sub Figure a Cantilever Zoom See 10 24 Engaging the cantilever on the sample Page 169 of 243...
Страница 170: ...Cypher SPM User Guide Sub Figure b Laser Spot ON See 11 Page 170 of 243 24 Engaging the cantilever on the sample...
Страница 173: ...Cypher SPM User Guide Sub Figure a Offset 24 Engaging the cantilever on the sample Page 173 of 243...
Страница 176: ...Cypher SPM User Guide Sub Figure a Master Panel Main Tab Page 176 of 243 24 Engaging the cantilever on the sample...
Страница 186: ...Cypher SPM User Guide Sub Figure a Before the Move Page 186 of 243 25 AC mode imaging...
Страница 189: ...Cypher SPM User Guide 26 26 Contact Mode Imaging In Air 26 Contact mode imaging in air Page 189 of 243...
Страница 216: ...Part VI Shipping and Maintenance This part of the user guide covers moving and shipping the instrument...