5. Always power off the microscope before exchanging scanners.
l
Before exchanging scanners, remember to first
motor the objective to its upper limit of travel.
l
Power off the controller before disconnecting the
scanner.
l
Exchange scanners and then power the controller
back on after the new scanner has been recon-
nected.
Note:
Pay attention! On older Cyphers, exchanging
the scanner while the system is powered up can
damage the backpack electronics.
6. Make sure the scanner is fully seated against the chassis before locking in place.
l
Unlike the S scanner, which has a hard stop on the
back of the scanner, the ES scanner has a spring
loaded stop that needs to be fully engaged.
l
When inserting the ES scanner into the chassis, push
the scanner until you feel the spring on the back of
the scanner engage. Fully depress the spring until
you feel the hard stop.
l
After fully engaging the spring loaded stop at the
back of the scanner, lock the scanner in place.
Note:
Proper seating of the scanner is most import-
ant when using the cooling capabilities of the
Cooler-Heater sample stage. During normal oper-
ation it is not particularly important.
7. Take care to properly set the correction collar on the objective.
l
The ES scanner uses different correction collar set-
tings than the S scanner.
l
The ES scanner uses 1.5 for gas operation and 2.0 for
liquid operation.
Note:
Incorrectly setting the correction collar will
degrade the optical image quality. In addition it will
cause slight errors in the XY calibration of the scan-
ner.
8.3 Chemical compatibility
The ES scanner is equipped with FKM (Viton equivalent) O-rings in the factory. For cases where
viton is chemically attacked, please switch to the included FFKM (Kalrez equivalent) O-ring. These
rings can be found in a small box:
8 Environmental Scanner Overview
Page 52 of 243
Содержание Cypher SPM
Страница 1: ...Cypher SPM User Guide Part No XX XXXX XXX...
Страница 8: ...Part I Getting Started System safety and overview...
Страница 165: ...Cypher SPM User Guide Sub Figure a First view of new cantilever 24 Engaging the cantilever on the sample Page 165 of 243...
Страница 167: ...Cypher SPM User Guide Sub Figure c Cantilever in focus See 9 24 Engaging the cantilever on the sample Page 167 of 243...
Страница 169: ...Cypher SPM User Guide Sub Figure a Cantilever Zoom See 10 24 Engaging the cantilever on the sample Page 169 of 243...
Страница 170: ...Cypher SPM User Guide Sub Figure b Laser Spot ON See 11 Page 170 of 243 24 Engaging the cantilever on the sample...
Страница 173: ...Cypher SPM User Guide Sub Figure a Offset 24 Engaging the cantilever on the sample Page 173 of 243...
Страница 176: ...Cypher SPM User Guide Sub Figure a Master Panel Main Tab Page 176 of 243 24 Engaging the cantilever on the sample...
Страница 186: ...Cypher SPM User Guide Sub Figure a Before the Move Page 186 of 243 25 AC mode imaging...
Страница 189: ...Cypher SPM User Guide 26 26 Contact Mode Imaging In Air 26 Contact mode imaging in air Page 189 of 243...
Страница 216: ...Part VI Shipping and Maintenance This part of the user guide covers moving and shipping the instrument...