3. Insert the replacement laser diode.
l
Push the laser diode module inwards with your fin-
ger with moderate force, while rotating the hex key
clockwise to tighten the new laser diode module, as
shown in the photograph to the right.
l
Replace the cover plate that was removed in step 1.
Tip:
Test that the new laser diode module works
properly before replacing the cover plate!
l
Replace the optics cover, if desired.
16.4.3 Focusing the blue laser spot
After installation of a blue laser module, the blue laser spot must be focused such that its focal
plane matches the focal plane of the infrared laser. The following procedure ensures the best focus.
1. Enable blueDrive.
2. Prepare for the blue laser focus.
l
Focus the objective onto the back of a cantilever chip
(the large body that you hold with tweezers when
handling the cantilever).
l
Position the infrared laser just underneath or above
a speck of dirt, which should be in sharp focus.
l
Move the blueDrive spot location such that it is in
line with the speck of dirt and the infrared laser, as
shown in the image to the right.
Page 131 of 243
16 blueDrive photothermal excitation
Содержание Cypher SPM
Страница 1: ...Cypher SPM User Guide Part No XX XXXX XXX...
Страница 8: ...Part I Getting Started System safety and overview...
Страница 165: ...Cypher SPM User Guide Sub Figure a First view of new cantilever 24 Engaging the cantilever on the sample Page 165 of 243...
Страница 167: ...Cypher SPM User Guide Sub Figure c Cantilever in focus See 9 24 Engaging the cantilever on the sample Page 167 of 243...
Страница 169: ...Cypher SPM User Guide Sub Figure a Cantilever Zoom See 10 24 Engaging the cantilever on the sample Page 169 of 243...
Страница 170: ...Cypher SPM User Guide Sub Figure b Laser Spot ON See 11 Page 170 of 243 24 Engaging the cantilever on the sample...
Страница 173: ...Cypher SPM User Guide Sub Figure a Offset 24 Engaging the cantilever on the sample Page 173 of 243...
Страница 176: ...Cypher SPM User Guide Sub Figure a Master Panel Main Tab Page 176 of 243 24 Engaging the cantilever on the sample...
Страница 186: ...Cypher SPM User Guide Sub Figure a Before the Move Page 186 of 243 25 AC mode imaging...
Страница 189: ...Cypher SPM User Guide 26 26 Contact Mode Imaging In Air 26 Contact mode imaging in air Page 189 of 243...
Страница 216: ...Part VI Shipping and Maintenance This part of the user guide covers moving and shipping the instrument...