Cypher SPM
User Guide
3. Mount your sample.
l
Use the low profile electrical sample puck assembly
(448.140) using silver paint or some other conductive
adhesive (upper image).
l
Unlike the S scanner, the ES scanner cannot accept
non-flat samples. A non-flat sample is any sample in
which the region to be scanned is not the tallest fea-
ture on the sample puck. For example, the sample
shown on the left of the lower image has a magnet
glued to the puck that is taller than the sample. The
low profile of the ES cantilever holders means the
magnet would hit the cantilever holder before the
cantilever engaged on the sample.
Attempting to starting a tip approach on a non-
flat sample may cause serious damage to your can-
tilever holder and sample stage.
4. Insert sample into the AFM.
l
Insert the sample puck as shown.
l
Use blunt tipped tweezers to connect the wire to the
front connection. This applies a sample bias.
l
Ensure the Engage Control Knob is fully coun-
terclockwise to facilitate sample access.
5. Follow steps 20–28 of
Loading the cantilever and sample on an environmental scanner on
You are now ready to begin imaging by controlling the instrument with the AZtecAFM software.
Page 194 of 243
28 Preparing for conductive AFM on the environmental scanner
Содержание Cypher SPM
Страница 1: ...Cypher SPM User Guide Part No XX XXXX XXX...
Страница 8: ...Part I Getting Started System safety and overview...
Страница 165: ...Cypher SPM User Guide Sub Figure a First view of new cantilever 24 Engaging the cantilever on the sample Page 165 of 243...
Страница 167: ...Cypher SPM User Guide Sub Figure c Cantilever in focus See 9 24 Engaging the cantilever on the sample Page 167 of 243...
Страница 169: ...Cypher SPM User Guide Sub Figure a Cantilever Zoom See 10 24 Engaging the cantilever on the sample Page 169 of 243...
Страница 170: ...Cypher SPM User Guide Sub Figure b Laser Spot ON See 11 Page 170 of 243 24 Engaging the cantilever on the sample...
Страница 173: ...Cypher SPM User Guide Sub Figure a Offset 24 Engaging the cantilever on the sample Page 173 of 243...
Страница 176: ...Cypher SPM User Guide Sub Figure a Master Panel Main Tab Page 176 of 243 24 Engaging the cantilever on the sample...
Страница 186: ...Cypher SPM User Guide Sub Figure a Before the Move Page 186 of 243 25 AC mode imaging...
Страница 189: ...Cypher SPM User Guide 26 26 Contact Mode Imaging In Air 26 Contact mode imaging in air Page 189 of 243...
Страница 216: ...Part VI Shipping and Maintenance This part of the user guide covers moving and shipping the instrument...