Cypher SPM
User Guide
3. Start imaging under the conditions (speed, scan size, and so on) you eventually plan to use on
the sample which was mounted and sealed inside the glovebox.
4. Observe image quality while you experiment with the inert gas flow and note the maximum
flow before imaging noise becomes noticeable.
5. Turn off the gas flow.
6. Stop imaging and disengage the tip.
7. Remove the cantilever holder.
8. Remove the entire sample cell from the AFM.
See
Exchanging the sample stage and cell body on page 84
.
12.2 Preparing the cell and gas lines
We will completely bypass the gas valve and pressure manifold for this demonstration.
1. Remove and store the gas lines that were already present on your Cypher AFM.
See
Gas handling and leak testing on page 93
2. Store the tubing sections in a safe place.
3. Consult the Cypher ES accessory parts list (see
Environmental scanner parts list on page 44
) and
find the following:
l
231.006 1/16” O.D. Tubing.
l
231.008 Luer Fittings: 231.008.
l
5cc Syringe: 080.010.
l
Threaded fittings: 232.015.
l
Scalpel knife with blade, 290.147 and 290.148
l
Ferrules: 232.016.
l
Extender Tool Fitting Wrench: 114.800.
l
Fitting Compression Fixture: 114.721
4. Locate a three way valve.
l
The luer-type valve shown is provided by Asylum.
5. Cut two sections of tubing of approximately 0.5 m
They should be long enough to reach from the scanner cell to the outside of the ES enclosure.
12 Glovebox Protocols
Page 98 of 243
Содержание Cypher SPM
Страница 1: ...Cypher SPM User Guide Part No XX XXXX XXX...
Страница 8: ...Part I Getting Started System safety and overview...
Страница 165: ...Cypher SPM User Guide Sub Figure a First view of new cantilever 24 Engaging the cantilever on the sample Page 165 of 243...
Страница 167: ...Cypher SPM User Guide Sub Figure c Cantilever in focus See 9 24 Engaging the cantilever on the sample Page 167 of 243...
Страница 169: ...Cypher SPM User Guide Sub Figure a Cantilever Zoom See 10 24 Engaging the cantilever on the sample Page 169 of 243...
Страница 170: ...Cypher SPM User Guide Sub Figure b Laser Spot ON See 11 Page 170 of 243 24 Engaging the cantilever on the sample...
Страница 173: ...Cypher SPM User Guide Sub Figure a Offset 24 Engaging the cantilever on the sample Page 173 of 243...
Страница 176: ...Cypher SPM User Guide Sub Figure a Master Panel Main Tab Page 176 of 243 24 Engaging the cantilever on the sample...
Страница 186: ...Cypher SPM User Guide Sub Figure a Before the Move Page 186 of 243 25 AC mode imaging...
Страница 189: ...Cypher SPM User Guide 26 26 Contact Mode Imaging In Air 26 Contact mode imaging in air Page 189 of 243...
Страница 216: ...Part VI Shipping and Maintenance This part of the user guide covers moving and shipping the instrument...