Cypher SPM
User Guide
1. Measure the Sample Bias voltage on the scanner's terminal block.
l
Check the 'Use' check box next to the Sample Voltage
parameter in the DoIV control panel.
Note:
In this example, the measured offset Bias
voltage is -46 mV.
2. Enter the amount of offset voltage needed in the S. Voltage Offset control parameter.
l
Use the opposite sign of the voltage measured on your
voltmeter to negate the actual voltage.
Note:
Make sure that the Sample voltage parameter is
set to 0 volts when making this adjustment.
3. Change the Sample voltage parameter and verify that the corresponding voltage appears on
the Sample pin on the scanner's terminal block.
27.2 Preparing the sample
Sample preparation varies but basically the goal is to provide an electrical path between the sample
bias and the surface of your sample. In addition to the electrical connection, care should be taken
to mount the sample mechanically to a sample puck as you would with any sample.
The ORCA kit comes with a practice sample of graphite (HOPG) mounted to a steel puck. A small
magnet is attached to the sample puck to provide an easy way for the bias lead to attach.
The magnetic connection method is convenient but is not necessary. A bias lead of your own
design can be mounted directly to the sample puck and used as long as the end of the lead is able
to fit into the sample voltage socket on the scanner's terminal block. Also, be certain to use wire
that is flexible enough not to impede normal scanning.
The following steps describe how this sample was prepared.
1. Use a small amount of 5 minute epoxy to attach the HOPG to the sample puck.
2. Place a magnet onto the puck.
3. Cover the sides of the sample and the entire magnet with silver paint.
Note:
The silver paint is not an adhesive. It will not provide good attachment of the sample
to the sample puck. Use the paint only to make an electrical connection from the sample to
the bias voltage lead.
27 Preparing for conductive AFM imaging on the standard scanner
Page 191 of 243
Содержание Cypher SPM
Страница 1: ...Cypher SPM User Guide Part No XX XXXX XXX...
Страница 8: ...Part I Getting Started System safety and overview...
Страница 165: ...Cypher SPM User Guide Sub Figure a First view of new cantilever 24 Engaging the cantilever on the sample Page 165 of 243...
Страница 167: ...Cypher SPM User Guide Sub Figure c Cantilever in focus See 9 24 Engaging the cantilever on the sample Page 167 of 243...
Страница 169: ...Cypher SPM User Guide Sub Figure a Cantilever Zoom See 10 24 Engaging the cantilever on the sample Page 169 of 243...
Страница 170: ...Cypher SPM User Guide Sub Figure b Laser Spot ON See 11 Page 170 of 243 24 Engaging the cantilever on the sample...
Страница 173: ...Cypher SPM User Guide Sub Figure a Offset 24 Engaging the cantilever on the sample Page 173 of 243...
Страница 176: ...Cypher SPM User Guide Sub Figure a Master Panel Main Tab Page 176 of 243 24 Engaging the cantilever on the sample...
Страница 186: ...Cypher SPM User Guide Sub Figure a Before the Move Page 186 of 243 25 AC mode imaging...
Страница 189: ...Cypher SPM User Guide 26 26 Contact Mode Imaging In Air 26 Contact mode imaging in air Page 189 of 243...
Страница 216: ...Part VI Shipping and Maintenance This part of the user guide covers moving and shipping the instrument...