15.2 Exchanging a light tube
You can exchange the Light tube without removing the top of the enclosure. However, we recom-
mended that you first familiarize yourself with the system by removing the top cover at least once
so that you are certain of the location of the components involved.
1. Remove your sample from the scanner.
2. Lower the cantilever holder.
l
Rotate the Engage Control Knob counter-clockwise
to lower the tip to a close distance (~1 mm) above
the top of the scanner. This enables easy access to
the top of the head assembly where the light tube is
located.
l
For safety reasons we recommend making motor
moves with the door closed.
Beware of pinch points. See
.
3. Turn the ARC2 controller power off.
This shuts off most of the instrument's power. You need not disconnect the motor power sup-
ply when exchanging the light tube assembly.
15 Light tubes
Page 116 of 243
Содержание Cypher SPM
Страница 1: ...Cypher SPM User Guide Part No XX XXXX XXX...
Страница 8: ...Part I Getting Started System safety and overview...
Страница 165: ...Cypher SPM User Guide Sub Figure a First view of new cantilever 24 Engaging the cantilever on the sample Page 165 of 243...
Страница 167: ...Cypher SPM User Guide Sub Figure c Cantilever in focus See 9 24 Engaging the cantilever on the sample Page 167 of 243...
Страница 169: ...Cypher SPM User Guide Sub Figure a Cantilever Zoom See 10 24 Engaging the cantilever on the sample Page 169 of 243...
Страница 170: ...Cypher SPM User Guide Sub Figure b Laser Spot ON See 11 Page 170 of 243 24 Engaging the cantilever on the sample...
Страница 173: ...Cypher SPM User Guide Sub Figure a Offset 24 Engaging the cantilever on the sample Page 173 of 243...
Страница 176: ...Cypher SPM User Guide Sub Figure a Master Panel Main Tab Page 176 of 243 24 Engaging the cantilever on the sample...
Страница 186: ...Cypher SPM User Guide Sub Figure a Before the Move Page 186 of 243 25 AC mode imaging...
Страница 189: ...Cypher SPM User Guide 26 26 Contact Mode Imaging In Air 26 Contact mode imaging in air Page 189 of 243...
Страница 216: ...Part VI Shipping and Maintenance This part of the user guide covers moving and shipping the instrument...