Cypher SPM
User Guide
3. Install the applications board.
l
Place the applications board over the connections on
the top right side of the Environmental scanner.
4. Seat the board.
l
Gently push the board onto the connectors.
5. Fit the cell body into the scanner and secure it to the engage stage ring with 3, 2-56 x 1/8”
screws.
6. Secure the stage base into the scanner.
Note:
The humidity cell is delivered with an Ambient Stage already installed. The humidity
cell body is compatible with all the Environmental Scanner Stages. If your experiment
requires temperature control then the ambient stage can be substituted for one of the act-
ive stages.
7. Connect the pigtail cable from the humidity cell body to the applications board.
l
Gently push the plug on the cable into the connector
on the board.
10 Cell Body and Sample Stage Guide
Page 72 of 243
Содержание Cypher SPM
Страница 1: ...Cypher SPM User Guide Part No XX XXXX XXX...
Страница 8: ...Part I Getting Started System safety and overview...
Страница 165: ...Cypher SPM User Guide Sub Figure a First view of new cantilever 24 Engaging the cantilever on the sample Page 165 of 243...
Страница 167: ...Cypher SPM User Guide Sub Figure c Cantilever in focus See 9 24 Engaging the cantilever on the sample Page 167 of 243...
Страница 169: ...Cypher SPM User Guide Sub Figure a Cantilever Zoom See 10 24 Engaging the cantilever on the sample Page 169 of 243...
Страница 170: ...Cypher SPM User Guide Sub Figure b Laser Spot ON See 11 Page 170 of 243 24 Engaging the cantilever on the sample...
Страница 173: ...Cypher SPM User Guide Sub Figure a Offset 24 Engaging the cantilever on the sample Page 173 of 243...
Страница 176: ...Cypher SPM User Guide Sub Figure a Master Panel Main Tab Page 176 of 243 24 Engaging the cantilever on the sample...
Страница 186: ...Cypher SPM User Guide Sub Figure a Before the Move Page 186 of 243 25 AC mode imaging...
Страница 189: ...Cypher SPM User Guide 26 26 Contact Mode Imaging In Air 26 Contact mode imaging in air Page 189 of 243...
Страница 216: ...Part VI Shipping and Maintenance This part of the user guide covers moving and shipping the instrument...