Cypher SPM
User Guide
3. Pull the cell body partway out.
l
Grasping the cell blocks, pull up on the cell body until
the diaphragm is fully extended.
4. Loosen fittings of gas inlet and outlet lines.
l
Grasp the cell body with your dominant hand.
l
With the opposite hand, use only your bare fingers
to loosen the fitting by turning it counterclockwise to
loosen it.
l
Detach both gas lines.
Note:
The gas lines are sealed by a small O-ring
(230.039) which may or may not come out when
you pull out the gas lines.
5. Check the gas lines to see if the small O-rings are stuck to the end.
Usually the rings stay inside the cell, but in case they are stuck to the gas lines, please unstick
them and place aside.
6. Lift out the sample stage.
l
Lift the sample stage out of the cell body.
l
If you encounter resistance, you may need to further
loosen the sample stage clamping screw (see step 1).
7. Store the sample stage / cell body combination it its proper storage container.
Refer to
Storing sample stage and cell body combinations on page 91
Page 87 of 243
10 Cell Body and Sample Stage Guide
Содержание Cypher SPM
Страница 1: ...Cypher SPM User Guide Part No XX XXXX XXX...
Страница 8: ...Part I Getting Started System safety and overview...
Страница 165: ...Cypher SPM User Guide Sub Figure a First view of new cantilever 24 Engaging the cantilever on the sample Page 165 of 243...
Страница 167: ...Cypher SPM User Guide Sub Figure c Cantilever in focus See 9 24 Engaging the cantilever on the sample Page 167 of 243...
Страница 169: ...Cypher SPM User Guide Sub Figure a Cantilever Zoom See 10 24 Engaging the cantilever on the sample Page 169 of 243...
Страница 170: ...Cypher SPM User Guide Sub Figure b Laser Spot ON See 11 Page 170 of 243 24 Engaging the cantilever on the sample...
Страница 173: ...Cypher SPM User Guide Sub Figure a Offset 24 Engaging the cantilever on the sample Page 173 of 243...
Страница 176: ...Cypher SPM User Guide Sub Figure a Master Panel Main Tab Page 176 of 243 24 Engaging the cantilever on the sample...
Страница 186: ...Cypher SPM User Guide Sub Figure a Before the Move Page 186 of 243 25 AC mode imaging...
Страница 189: ...Cypher SPM User Guide 26 26 Contact Mode Imaging In Air 26 Contact mode imaging in air Page 189 of 243...
Страница 216: ...Part VI Shipping and Maintenance This part of the user guide covers moving and shipping the instrument...