Cypher SPM
User Guide
1. Place the cell onto the scanner.
l
Seat the foot of the cell onto the receptacle located
on top of the scanner.
l
The trapped air inside the cell will prevent the tip
from colliding with the sample.
2. Lock the foot.
l
Lock the foot using the tool as shown.
l
Do not over tighten. Use only your fingertips to
handle the tool.
3. Press down and seat the cell body.
l
Press down on the sample stage as shown. You will
see the membrane bulge a bit under the building
pressure. This pressure will assist the proper forming
of the membrane.
l
During this process, retract on the syringe to relieve
the pressure as the cell lowers.
l
About 3 cm
3
of gas should be removed to properly
seat the cell body.
Tip:
A little overpressure is better than a slight
vacuum in terms of keeping contaminants out of
the cell.
Page 103 of 243
12 Glovebox Protocols
Содержание Cypher SPM
Страница 1: ...Cypher SPM User Guide Part No XX XXXX XXX...
Страница 8: ...Part I Getting Started System safety and overview...
Страница 165: ...Cypher SPM User Guide Sub Figure a First view of new cantilever 24 Engaging the cantilever on the sample Page 165 of 243...
Страница 167: ...Cypher SPM User Guide Sub Figure c Cantilever in focus See 9 24 Engaging the cantilever on the sample Page 167 of 243...
Страница 169: ...Cypher SPM User Guide Sub Figure a Cantilever Zoom See 10 24 Engaging the cantilever on the sample Page 169 of 243...
Страница 170: ...Cypher SPM User Guide Sub Figure b Laser Spot ON See 11 Page 170 of 243 24 Engaging the cantilever on the sample...
Страница 173: ...Cypher SPM User Guide Sub Figure a Offset 24 Engaging the cantilever on the sample Page 173 of 243...
Страница 176: ...Cypher SPM User Guide Sub Figure a Master Panel Main Tab Page 176 of 243 24 Engaging the cantilever on the sample...
Страница 186: ...Cypher SPM User Guide Sub Figure a Before the Move Page 186 of 243 25 AC mode imaging...
Страница 189: ...Cypher SPM User Guide 26 26 Contact Mode Imaging In Air 26 Contact mode imaging in air Page 189 of 243...
Страница 216: ...Part VI Shipping and Maintenance This part of the user guide covers moving and shipping the instrument...