Temperature Input Unit Functions
Section 11-5
8.
Adjust the lower limit value. Input 0 mV from the reference voltage/current
generator (STV) to the Temperature Input Unit's input terminals. Wait at
least 1 minute for the input to stabilize.
9.
Click the
Fix lower adjusting Value
Button in the Adjustment Window to
write the adjustment value.
10. Adjust the upper limit value. Input the upper limit voltage from the reference
voltage/current generator to the input terminals of the input to be adjusted.
Refer to the following table for the appropriate voltage. Wait at least 1
minute for the input to stabilize.
11. Click the
Fix upper adjusting Value
Button and write the adjustment val-
ue.
Type
Input voltage
K (
−
200 to 1300
°
C)
50 mV
K (0.0 to 500.0
°
C)
20 mV
J (
−
100 to 850
°
C)
50 mV
J (0.0 to 400.0
°
C)
20 mV
T
20 mV
L (
−
100 to 850
°
C)
50 mV
L (0.0 to 400.0
°
C)
20 mV
U
20 mV
N
50 mV
PL2
50 mV
Содержание CRT1 - 10-2008
Страница 2: ...CRT1 Series CompoNet Slave Units and Repeater Unit Operation Manual Revised October 2008...
Страница 3: ...iv...
Страница 5: ...vi...
Страница 9: ...x...
Страница 15: ...xvi...
Страница 23: ...xxiv Conformance to EC Directives 6...
Страница 51: ...28 Slave Unit Models Section 1 2...
Страница 123: ...100 Connecting External I O for Slave Units Section 3 6...
Страница 365: ...342 Temperature Input Units Section 7 5...
Страница 381: ...358 Expansion Unit Specifications Section 8 2...
Страница 418: ...395 Repeater Unit Section 10 2 Dimensions 42 3 50 3 1 27 5 95 mm...
Страница 419: ...396 Repeater Unit Section 10 2...
Страница 424: ...401 CX Integrator Section 11 1...
Страница 575: ...552 Current Consumption Summary Appendix D 552...
Страница 581: ...558 Index...
Страница 583: ...560 Revision History...