DS2 Docking Station
Troubleshooting
Version 9.1 (P/N: 17112798)
INDUSTRIAL SCIENTIFIC
249
To access the Event Log on another computer, follow the instructions that are listed below.
Step Instruction
1.
Click on the Start button. If you are using Windows
2000, select Settings.
2.
Select Control Panel. The Control Panel window opens.
3.
Double-click on the Administrative Tools icon. The Administrative Tools window
opens.
4.
Double-click on the Computer Manager icon. The Computer Manager utility opens.
5.
Click on the Action menu and choose Connect to another computer. Select the server
whose Event Log you wish to view.
6.
Once you have connected to remote computer, click on Event Viewer in the left pane
of the window. Please refer to your operating system‘s user guide for detailed
information about using the Event Viewer.
Table 14-1. Sample Event Log Messages
Error
Condition
Event Viewer Message
Required gas is
not connected
(no iGas
connected and
no gas manually
configured
through DSSAC)
******* EVENT VIEWER MESSAGE: Warning 0 Event type:
InstrumentCalibrationForcedEvent
Docking Station: 0403525-075
Docked Instrument: 0211350084
Description: Resources for event on 0403525-075 were
unavailable.Could not find cylinder needed for sensor:
001714522082#S0001
Error Text:
*******END EVENT VIEWER MESSAGE*******
Required gas is
connected, but
expired
******* EVENT VIEWER MESSAGE: Warning 0 Event type:
InstrumentCalibrationForcedEvent
Docking Station: 0403525-075
Docked Instrument: 0211350084
Description: Resources for event on 0403525-075 were
unavailable.Cylinder has expired.
Cylinder id =83075A-104
Port number: 2
Could not find cylinder needed for sensor:
001714522082#S0001
Error Text:
*******END EVENT VIEWER MESSAGE*******