Configuring the DS2
DS2 Docking Station
190
INDUSTRIAL SCIENTIFIC
Version 9.1 (P/N: 17112798)
To change gas cylinders on an IDS, follow the instructions listed below.
Step Instruction
1.
Open the DSSAC application. Click the Docking Stations option in the navigation
pane.
2.
Double-click on the IDS for which you are changing gas cylinders.
Alternative:
You can also right-click on the IDS and select Properties from the
context menu.
Alternative:
You can also select the IDS, and then click the File menu, and select
Properties.
3.
Click on the Gas Inlets tab.
4.
Click on the Gas In tab that corresponds to the Gas In connection for which you are
changing gas cylinders.
5.
Click the Assign button, and select Disabled from the menu. This disables the Gas In
connection.
6.
Click OK from the Edit Docking Station dialog box. Click Yes when the Confirm
Docking Station Save message appears.
7.
Disconnect the old gas bottle from the IDS. Remove the gas tubing from the fitting
on the demand flow regulator. With the gauge facing away from you, unscrew the
gas cylinder bottle until it detaches from the regulator.
8.
Connect the new gas bottle to the IDS. See section 6.6 Configuring Gas Cylinders
for detailed instructions on connecting gas cylinders.
9.
Return to the DSSAC application. Click the Docking Stations option in the
navigation pane.
10.
Double-click on the IDS for which you are changing gas cylinders.
11.
Click on the Gas Inlets tab.
12.
Click on the Gas In tab that corresponds to the Gas In connection for which you are
changing gas cylinders.
13.
Click the Assign button, and select ISC or non-ISC from the menu, depending on the
gas cylinder you are adding. See the previous sections on using ISC gas or using
non-ISC gas for specific procedures for adding ISC or non-ISC cylinders.
14.
Click OK from the Edit Docking Station dialog box.
15.
The Gas In connection is now configured to use the new gas cylinder.