DS2 Docking Station
Configuring the DS2
Version 9.1 (P/N: 17112798)
INDUSTRIAL SCIENTIFIC
185
Step Instruction
8.
Click the Assign button. A menu of options appears. Choose from the options
below.
Fresh Air – Configures the Gas In connection to use fresh ambient air, or for a
Zero Air cylinder.
ISC – Configures the Gas In connection to use an Industrial Scientific gas
cylinder. See the Using ISC gas section for specific procedures for adding
ISC gas.
Non-ISC – Configures the Gas In connection to use a gas cylinder that was
not purchased from Industrial Scientific. See the Using non-ISC gas for the
specific procedures for adding non-ISC gas.
Disabled – Configures the Gas In connection to be disabled, i.e., not using
either a gas cylinder or fresh air.
9.
The settings you have chosen appear in the dialog box on the Gas In tab that you had
selected. If the Gas In connection is configured with gas, then the Gas Name,
Concentration and Gas Symbol displays on the screen. The expiration date of the gas
cylinder appears above the list of gases.
Figure 6-12. A Configured Gas In Tab